70 Dako Autostainer Plus | User Guide
5. Enter the correct compatibility code and press TAB or ENTER. Assigning compatibility
information to reagents is optional (See Compatibility Check at the end of this section).
The cursor moves to the Lot box.
6. Enter the reagent lot number and press TAB or ENTER. The lot # cannot exceed 8 alpha-
numeric characters. Assigning lot number information to reagents is optional. The cursor
moves to the Exp. Date box.
7. Enter the reagent expiration date and press TAB or ENTER. The reagent expiration date
entry is optional. To enter, use XXYY where XX is a two-digit numeric character
symbolizing a month and YY signifies a year. A maximum of 5 characters can be entered.
The entry of expired reagents is not accepted by the software. The cursor moves to the
Time box.
NOTE: All expiration dates entered are limited to a maximum of 10 years from the date of
entry.
8. Enter the reagent incubation time and press TAB or ENTER. The cursor moves the
Reagent Name box to allow addition of the next reagent within the same reagent
category. Follow steps 3-8 until all reagents have been added.
NOTE: A minimum of 2 minutes incubation time is required for proper operating
conditions.
9. Select the OK button to add the new reagent(s) to the reagent list. The system also
prompts “Save changes on disk?”. Select the YES button and the new reagent is saved to
the appropriate list. Select the NO button and the new reagent is not saved. The
PROGRAMMING GRID is displayed.
Select the CANCEL button to escape without saving. The PROGRAMMING GRID is
displayed.
Adding Primary Antibodies
A primary antibody can be added to the reagent list with a specific pretreatment reagent. This will
facilitate automatic selection of the pretreatment whenever a primary antibody is selected at the
PROGRAMMING GRID.
1. From the PROGRAMMING GRID select Edit lists from the menu bar. The reagent lists
categories are displayed.
2. Select the Primary Antibody category. The Primary Antibody EDIT REAGENT LIST
screen is displayed with the cursor in the text box in the Antibody section.
NOTE: All symbol characters such as (*), (:), (/) and (-), etc. are incompatible with the
Autostainer Plus software. Do not use any symbol characters. The use of symbol
characters may interrupt the Autostainer Plus during operation.